The IATROSCAN MK-7s is thin layer chromatograph measuring sample(s) by means of Flame Ionization Detector(FID). The sample developed on a thin layer quarts rod 'Chromarod' designed for IATROSCAN's exclusive use can be detected by IATROSCAN automatically and quickly. Neither troublesome coloring nor activation of thin layer for hours is required for IATROSCAN analysis. Direct detection can be made for a large variety of organic compounds, including not only those analyzable by TLC but also samples of a higher boiling point which are difficult to analyze by GC as well as those problematic to detect by LC.
Heavy oil by IATROSCAN (TLC-FID/FPD)
Sample:Bitumem
- Saturated H.C.
- Aromatic
- Resin
- Asphaltene
Stationary phase :Chromarod-SIII
Mobile phase :
1st. Hexane 100% 10cm
2nd. Toluene 100% 6cm
3rd. Dichloromethane:Methanol 57:3 2.5cm
Lipids by IATROSCAN (TLC-FID/FPD)
Sample:Standard mixture
- Cholesterol ester
- Triglyceride
- Cholesterol
- Phosphatidyl ethanolamine
- Phosphatidyl choline
- Phosphatidyl inositol
- Sphingomyelin
- Lysophosphatidyl choline
Stationary phase :Chromarod-SIII
Mobile phase :
1st. Chloroform:Methanol:Water: 25%Ammonia
47:20:2.5:0.28 7cm
2nd.Hexane :Diethyl ether 63:7 10cm
Glycerides by IATROSCAN (TLC-FID)
Sample:Standard mixture
- TG (tripalmitin)
- FA (palmitic acid)
- 1,3-DG (1,3-dimyristin)
- 1,2-DG (1,2-dipalmitin)
- MG (monopalmitin)
Stationary phase : CHROMAROD-SIII
Mobile phase :
Benzene :Chloroform : Acetic acid
50 : 20 : 0.7 10cm